2025 SMSI Bannerklein

P2.1.20 Single-beam multi-cantilever optical measurement head for cantilever array-based biosensors

Event
14th International Meeting on Chemical Sensors - IMCS 2012
2012-05-20 - 2012-05-23
Nürnberg/Nuremberg, Germany
Chapter
P2.1 Biosensors
Author(s)
K. Nieradka, D. Kropiec, T. Gotszalk - Faculty of Microsystem Electronics and Photonics, Wrocław University of Technology (Poland), P. Grabiec, M. Wegrzecki - Division of Silicon Microsystem and Nanostructure Technology, Institute of Electron Technology (Poland)
Pages
1386 - 1389
DOI
10.5162/IMCS2012/P2.1.20
ISBN
978-3-9813484-2-2
Price
free

Abstract

Here we present a homemade system for highly sensitive measurements of cantilever sensor arrays. Two unique techniques developed by our group are utilized. Firstly, the expanded beam deflection (EBD) method that allows truly simultaneous measurements of multiple cantilevers with a simple optical apparatus consisting of a single and stationary laser beam, while providing the same sensitivity as the original optical beam deflection (OBD) method. Secondly, the real-time Brownian noise extraction technique allows combined static-dynamic mode operation in gaseous or liquid environment with no need for external excitation of vibrations. Thus constructed measurement system allows highly sensitive simultaneous measurement of bending as well as flexural and torsional vibrations of up to 8 cantilevers.

Download